Understanding Hemocompatibility in Medical Devices

Medical devices that come into contact with blood must be engineered for hemocompatibility — the ability to avoid triggering the body’s clotting and immune response. Poor hemocompatibility can result in thrombus formation on device surfaces, impairing function and creating downstream risks such as emboli, blocked vessels, and tissue damage.

Why Hemocompatibility Matters

Erythrocytes (red blood cells) are trapped within a fibrin mesh.
Fig.1

Blood readily reacts to foreign materials. When exposed to a non-hemocompatible surface, platelets activate, fibrin forms, and clots begin to develop. These can:

  • Disrupt the function of implanted or extracorporeal devicesCompromise patient safetyLead to device failure or systemic complications

For high-risk applications like neurovascular implants, venous stents, or oxygenators in ECMO circuits, hemocompatibility is not optional — it’s foundational.

How Clots Form: Coagulation Pathways

red blood cells,activated platelet and white blood cells microscopic photos
Fig.2

When a device enters the bloodstream, it activates the intrinsic coagulation pathway. This is different from the extrinsic pathway, which is triggered by tissue injury. Key steps include:

  • Plasma protein adsorption (e.g., fibrinogen, albumin — the Vroman effect)
  • Thrombin generation – a pivotal enzyme that converts fibrinogen to fibrin
  • Fibrin mesh formation, trapping red cells and platelets
  • Platelet activation, further amplifying the coagulation process

In parallel, white blood cells like neutrophils may release NETs (neutrophil extracellular traps), providing scaffolds for additional clotting activity.

Design Factors That Influence Hemocompatibility

Hemocompatibility is not just a material issue — it’s influenced by multiple device design variables:

Design ElementImpact on Hemocompatibility
Material surface chemistryAffects protein and platelet interactions
Geometry & flow dynamicsCan cause stagnation or turbulence that promotes clotting
Surface area and roughnessLarger or rougher areas increase activation risk
Duration of contactLong-term implants have stricter requirements

Microscale flow stagnation zones or areas of high shear can both elevate clot risk. Poorly optimized surfaces also raise the chance of hemolysis and immune activation.

The medical device design process needs to consider evaluating a device’s hemocompatibility at an early stage to ensure its ability to perform in blood is “designed in.” Various in-vitro and in-vivo tests have been developed that consider the complexity of the clotting cascade and the device’s surface/blood interactions.

Blood drawn from a patient with Serum separate in the chemistry laboratory. Lab technician holding a test tube of blood sample after being centrifuged
Fig.4

The most common in-vitro assays rely on either static or dynamic blood contact. In-vitro dynamic testing produces the most detailed laboratory testing for evaluating a device’s performance without the cost, complexity, and ethical considerations of animal testing.

 This testing can also be performed on individual materials and device sub-systems at a time when the entire device may not be ready for in-vivo evaluation, allowing developers to get an early indicator of device performance. 

The table below shows some of the common test methods that are used and which cover the performance of a device with respect to the different mechanisms of clot initiation.

CategoryTestPurpose
CoagulationMeasurement of Thrombin-Antithrombin (TAT) complex using ELISA measurementThe purpose of the TAT Elisa assay is to determine the progression of sample blood towards coagulation as an indicator of the formation of fibrin networks.
HemolysisThe measurement of Free Hemoglobin by spectrophotometryThis measurement provides an indicator for how many erythrocytes are being lysed (freeing up hemoglobin). This tells how likely the device surface is to damage blood and lead to pro-coagulant constituents being released.
HematologyMeasurement of circulating cell counts (Platelets, White Blood Cells, and Red Blood Cells) using commercial hematology analyzersReduction in circulating cell counts suggest that cells have either adsorbed on the surface or otherwise been incorporated into a blood clot.
PlateletMeasurement of markers of platelet activation such as Beta-Thromboglobulin or Platelet Factor 4 using an ELISA AssayThe biological markers indicating platelet activation
ComplementMeasurement of complement activation markers such as C3a, C5a and SC5b9 using an ELISA assayThe activation of the complement system indicates an innate immune response to the surface of the device which can lead to coagulation
Leukocyte ActivationMeasurement of PMN Elastase using an ELISA AssayThe activation of PMN elastase indicates that there is an immune response mediated by leukocytes.
Surface analysisMicroscopic Surface evaluation using SEM or light microscopyThe visual appearance of surfaces, particularly under SEM, can identify challenge areas and adherence of fibrin, platelets, and blood cells

Dynamic testing systems simulate realistic flow conditions and can be applied early in the design phase, before full prototypes are developed.

Not all device materials are naturally blood-compatible. Materials like polypropylene or PMP often require surface enhancements to meet safety thresholds.

Camouflage™, Smart Reactors’ proprietary coating, helps solve this problem. It offers:

  • Drug-free thromboresistance
  • Reduced platelet adhesion
  • Support for endothelialization

By modifying the device surface without altering mechanical performance, Camouflage™ enhances both short-term patency and long-term healing.

Thrombus formation on blood-contacting medical devices is a preventable complication. Manufacturers must proactively design for hemocompatibility by considering:

  • Material selection and surface chemistry
  • Flow dynamics and geometry
  • Surface coatings that reduce biological reactivity

At Smart Reactors, we support manufacturers through every phase of development — from prototype coatings to clinical-grade production — helping ensure their devices perform safely inside the body.